FMS 102 probes TESA, ± 2 mm, measuring travel 5.8 mm, radial cable
This product is priced individually. Fill in the form to receive a dedicated quote.
Informazioni di prodotto e di utilizzo
TESA FMS 102 inductive probe, measuring range ± 2 mm.
Description
FMS102 probe, ± 2 mm, radial cable.
Probe:
- Linear guide on a ball bearing
- Compact and robust construction
- Protected probes shield the 2 side surfaces against dust
Measurement:
- Parallel guide
- Can be mounted in any position
- Adjustable measuring direction
- Adjustable retraction direction of the contact point
- Variable measuring force
- Offset contact points can be used
- Contact points can be replaced or changed
TESA FMS 102 inductive probe, compatible with the M4P-3 4-input interface with analogue output.
Technical data. Measuring range ± 2 mm; nominal output signal with M4P-3 100 mV for a 100 µm travel from 0 with 1 V/mm gain.
Key features
- Linear guide on a ball bearing
- Compact and robust construction
- Parallel guide
- Can be mounted in any position
- Adjustable measuring direction
- Measuring range ± 2 mm
Technical data
| Series | FMS102 |
| Measuring range [mm] | ± 2 |
| Repeatability limit [µm] | 0.5 |
| Hysteresis [µm] | 0.5 |
| Maximum permissible linearity error [µm] | 0.2 + 3 x (L x L x L) |
| Measuring force [N] | 2 |
| Plunger lift | Retraction by pneumatic pressure (optional accessory) |
| Protective bellows | None |
| Cable orientation | Radial |
| Sensitivity [mV/V/mm] | 73.75 |
| Cable length [m] | 2 |
| Standard | DIN 32876 |
| Material | Probe body: nickel-plated steel |
| Dimensions | Cable: 2 m; Connector: 5p DIN 45322 |
| Mass | Moving mass: 110 g |
| Hysteresis | 0.5 µm |
| Coefficient of linear expansion | 0.14 µm/°C |
| Repeatability | 0.5 µm |
| Measuring range | ± 2 mm |
| Plunger travel | 5.8 mm |
| Working temperature | 20 °C ± 0.5 |
| Sensitivity | 73.75 mV/V/mm |
| Mounting of the measuring instrument | 4x M6 mounting threads |
| Maximum permissible linearity error | L in mm; 0.2 + 3 x (L x L x L) µm |
Models and variants
| TESA code | Item | Protection rating | Mass [kg] | Mass [g] |
|---|---|---|---|---|
| 03230028 | FMS 102 parallel-guided probe / F= 2 N / Measuring range: ±2 | IP50 | 0.429 | 429 |
| 03230038 | FMS 102-P parallel-guided probe / F= 2 N / Measuring range: ±2 / Protected version | IP54 | 0.399 | 399 |
Supplied with
Probe · Hex key · 2x M2 x 10 screws · 2x M6 x 20 screws · 2x M6 washers · Inspection report · Instruction manual · Declaration of conformity
Documentation
- Document: 15-08-32-catalog extract-fms102 probes 2 mm 5 8 mm measuring trav
- Document: 15-08-32-catalog extract-fms probes usb dc axial probes neutral a
- Document: 15-08-32-catalog extract-numerical index index 3 cat gen en p481
- Data sheet / dimensions: 03200254a
- Data sheet / dimensions: 03200289a
Official manufacturer page: tesatechnology.com
Price and availability
Products without a listed price are offered in a configuration defined together with our technical support: contact us for a quote stating the model, quantity and application.
Selection guide: TESA FMS 102 probes, ± 2 mm, measuring travel 5.8 mm, radial cable
TESA FMS 102 inductive probe, measuring range ± 2 mm.
TESA inductive probes (half-bridge system) are selected according to measuring range (± 1, ± 2, ± 5 mm), measuring force (from 0.16 to 4 N depending on the material and finish of the workpiece), cable exit (axial or radial, depending on the space available in the fixture) and plunger lift (manual or pneumatic, GTx2 and GTLx2 versions). The GTL versions are linearised for maximum accuracy over the whole range; the DC versions have a voltage analogue output for PLCs; the USB versions connect directly to a PC without a display unit; the FMS parallel-guided probes are for measurements with constant force and interchangeable probes; lever probes are for hard-to-reach points. The probe must be paired with a display unit (TWIN, TT100) or an interface (M4P-3, TESA-MODUL) and with suitable contact points.
If in doubt, contact us: our technical support will advise you on the right model, measuring force, contact points and display unit for your application.
Integration and programming
The probe is fixed by its Ø 8 mm h6 shank in the holder or measuring fixture and connected to the display unit or interface via the 5-pin DIN connector (or USB for the USB versions, voltage output for the DC versions). Zeroing, presets, tolerances and formulas between several probes are set on the display unit; with M4P-3 and TESA-MODUL the signals go to PLCs, acquisition boards and SPC software. Probes with pneumatic plunger lift require the pneumatic accessories (electropneumatic pump, fittings) listed among the accessories.
Calibration and repair
Every TESA probe is supplied with an inspection report and a declaration of conformity. Periodic verification on gauge blocks or setting standards is recommended and, for quality systems, calibration with a traceable certificate; replace worn contact points and damaged bellows. After impacts or unstable readings the probe should be sent in for service.
Service: RODER S.r.l. – INNOVACHECK division handles verification, calibration with certificate and repairs of TESA probes and display units in Italy. Contact us stating the model and serial number.